AR6 portable network analyzer
User Manual Page 23 of 69
8.1.2 Time to switch off display
This field is to select the time during which the screen will light after the last change. Once this
time has elapsed, the screen will switch OFF and go into a low power status to increase battery
autonomy and life. In this status, the Power ON/OFF button will be flashing in green.
8.1.3 Contrast
This field is to select the screen contrast. This field is given in %. A correct configuration can
extend battery autonomy and life.
8.1.4 Time to lock keyboard
This field is to select Enter the desired time, in minutes, after which the button pad will lock. If
this function is not desired, enter a zero (0) value.
8.1.5 Password
This field is to select a password, if used. Enter a 4-figure password which will be used to
unlock the unit's button pad once the locking time has elapsed.
8.1.6 Phase colours
The AR6 has colour templates to identify the phases. The available names and colours are
shown in the following table:
Table 8-1.- Options of colour codes to identify phases
L1 / A L2 / B L3 / C LN / N LK / K earth
EU black red yellow blue grey green
UK red yellow blue black grey green
CA red black
white grey green
USA black red
white grey green
8.1.7 Screen background
The user can change the screen background using this option of the menu. The selected image
will be used as background for all screens.
8.2 CLOCK SETUP
Fig. 8-2.- Clock configuration
Displayed parameters:
The “Clock setup” options allow the
configuration of the date and time format and
the desired date and time.
It also allows the selection of time zone.
WARNING !Changing the unit time will cause all the data logged on the memory
to be erased
Commentaires sur ces manuels